Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Fluke 87: (1 year) CAL VER /5720,5725 DATE: 10-Sep-98 AUTHOR: Fluke Corporation REVISION: $Revision: 1.2 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 71 NUMBER OF LINES: 265 CONFIGURATION: Fluke 5720A CONFIGURATION: Fluke 5725A ============================================================================= # # Source: # Fluke 80 Series Service Manual. # (PN 834168 April 1989, Rev. 3, 7/93, Change/Errata 8/94). # # Compatibility: # MET/CAL 4.1 or later # # Subprocedures: # Sub Fluke 83,85,87 (1 year) Capacitance CAL VER # Sub Fluke 83,85,87 (1 year) Frequency CAL VER /3325 # # Required Files: # None # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5720A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # # The 90 day specifications of the 5720A are used in TUR computations. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- R N P F W 1.002 ASK+ K 1.003 HEAD PRELIMINARY INSTRUCTIONS 1.004 DISP [32] WARNING 1.004 DISP HIGH VOLTAGE is used or exposed during the performance 1.004 DISP of this calibration. DEATH ON CONTACT may result if 1.004 DISP personnel fail to observe safety precautions. 1.005 RSLT =DISPLAY TEST 1.006 PICE 87_dsp 2.001 DISP No warm-up time requirement. 2.001 DISP Ambient temperature: 18C - 28C. 2.001 DISP Relative humidity: less than 90%. 2.002 ASK- U 2.003 HEAD {ROTARY SWITCH TESTS} 2.004 DISP Hold down the RANGE button while turning the UUT rotary 2.004 DISP function switch from OFF to VDC. Release the RANGE 2.004 DISP button and turn the switch to {VAC}. 2.005 MATH MEM1 = 0 2.006 MEMI Enter UUT reading: 2.007 MEME 2.008 MEMC display 12U 3.001 DISP Rotate the UUT function switch to {VDC}. 3.002 MATH MEM1 = -32 3.003 MEMI Enter UUT reading: 3.004 MEME 3.005 MEMC display 12U 4.001 DISP Rotate the UUT function switch to {mVDC}. 4.002 MATH MEM1 = -64 4.003 MEMI Enter UUT reading: 4.004 MEME 4.005 MEMC display 12U 5.001 DISP Rotate the UUT function switch to {Ohms}. 5.002 MATH MEM1 = -96 5.003 MEMI Enter UUT reading: 5.004 MEME 5.005 MEMC display 12U 6.001 DISP Rotate the UUT function switch to {Diode}. 6.002 MATH MEM1 = -128 6.003 MEMI Enter UUT reading: 6.004 MEME 6.005 MEMC display 12U 7.001 DISP Rotate the UUT function switch to {mA/A}. 7.002 MATH MEM1 = -160 7.003 MEMI Enter UUT reading: 7.004 MEME 7.005 MEMC display 12U 8.001 DISP Rotate the UUT function switch to {uA}. 8.002 MATH MEM1 = -192 8.003 MEMI Enter UUT reading: 8.004 MEME 8.005 MEMC display 12U 9.001 ASK+ U 9.002 HEAD {AC VOLTAGE TESTS} 9.003 DISP Rotate the UUT function switch from OFF to VAC. 9.004 DISP Connect the 5720A and the UUT as follows. 9.004 DISP [32] 5720A OUTPUT HI to UUT V Ohm DIODE 9.004 DISP [32] 5720A OUTPUT LO to UUT COM 9.004 DISP [32] 5720A SENSE HI to UUT V Ohm DIODE 9.004 DISP [32] 5720A SENSE LO to UUT COM 9.005 HEAD AC VOLTAGE TESTS: {400mV Range} 9.006 5720 400 350.0mV 2.9U 60H 2W 10.001 5720 400 350.0mV 3.9U 1kH 2W 11.001 5720 400 350.0mV 7.4U 5kH 2W 12.001 5720 400 350.0mV 9.0U 20kH 2W 13.001 HEAD AC VOLTAGE TESTS: {4V Range} 13.002 5720 4 3.500V 0.027U 60H 2W 14.001 5720 4 3.500V 0.039U 1kH 2W 15.001 5720 4 3.500V 0.074U 5kH 2W 16.001 5720 4 3.500V 0.090U 20kH 2W 17.001 HEAD AC VOLTAGE TESTS: {40V Range} 17.002 5720 40 35.00V 0.27U 60H 2W 18.001 5720 40 35.00V 0.39U 1kH 2W 19.001 5720 40 35.00V 0.74U 5kH 2W 20.001 5720 40 35.00V 0.90U 20kH 2W 21.001 HEAD AC VOLTAGE TESTS: {400V Range} 21.002 5720 400 350.0V 2.7U 60H B1 2W 22.001 5720 400 350.0V 3.9U 1kH B1 2W 23.001 5720 400 350.0V 7.4U 5kH B1 2W 24.001 5720 400 100.0V 4.0U 20kH 2W 25.001 5720 400 200.0V 6.0U 20kH 2W 26.001 5720 400 300.0V 8.0U 20kH B1 2W 27.001 5720 400 350.0V 9.0U 20kH B1 2W 28.001 HEAD AC VOLTAGE TESTS: {1000V Range} 28.002 5720 1000 900V 8U 60H B1 2W 29.001 5720 1000 900V 13U 1kH B1 2W 30.001 5720 1000 900V 22U 5kH B1 2W # Better TURs for the frequency tests can be achieved using a HP 3325B # function generator. If you wish to use the function generator in lieu # of the 5700A delete the 5700A based frequency tests and uncomment the # CALL statement at the bottom of this procedure. 31.001 HEAD {FREQUENCY TESTS} 31.002 DISP Press the RANGE button to lock the 400mV Range. 31.002 DISP Press the Hz button to choose the Frequency mode. 31.003 HEAD FREQUENCY TESTS: {19.999kHz Range} 31.004 5720 20 19.000kH 0.002U 150mV 2W #! WARNING: Test Tol 2, Sys Tol 1.9, TUR 1.0526 (< 4.00). 32.001 HEAD FREQUENCY TESTS: {199.99kHz Range} 32.002 5720 200 190.00kH 0.02U 150mV 2W #! WARNING: Test Tol 20, Sys Tol 19, TUR 1.0526 (< 4.00). 33.001 HEAD {FREQUENCY SENSITIVITY AND TRIGGER LEVEL} 33.002 HEAD FREQUENCY SENSITIVITY AND TRIGGER LEVEL: {4V AC Range} 33.003 DISP Press the RANGE button to lock the 4V Range. 33.004 5720 1000.0H 0.2U 300mV 2W #! WARNING: Test Tol 0.2, Sys Tol 0.1, TUR 2.0000 (< 4.00). 34.001 HEAD FREQUENCY SENSITIVITY AND TRIGGER LEVEL: {4V DC Range} 34.002 DISP Rotate the UUT function switch to VDC. 34.002 DISP Press the RANGE button to lock the 4V Range. 34.002 DISP Press the Hz button to choose the Frequency mode. 34.003 5720 1000.0H 0.2U 1.7V 2W #! WARNING: Test Tol 0.2, Sys Tol 0.1, TUR 2.0000 (< 4.00). 35.001 5720 1000.0H 1.0V S 2W 35.002 EVAL Is the UUT {displaying all zeros}? 36.001 HEAD FREQUENCY SENSITIVITY AND TRIGGER LEVEL: {40V DC Range} 36.002 DISP Press the RANGE button to lock the 40V Range. 36.003 5720 1000.0H 0.2U 6.0V 2W #! WARNING: Test Tol 0.2, Sys Tol 0.1, TUR 2.0000 (< 4.00). 37.001 5720 1000.0H 2.0V S 2W 37.002 EVAL Is the UUT {displaying all zeros}? 38.001 HEAD {DC VOLTAGE TESTS} 38.002 DISP Press the Hz button twice to select VDC. 38.002 DISP Press the RANGE for 2s to select autorange. 38.003 HEAD DC VOLTAGE TESTS: {4V Range} 38.004 5720 4 3.500V 0.005U 2W 39.001 HEAD DC VOLTAGE TESTS: {40V Range} 39.002 5720 40 35.00V 0.05U 2W 40.001 5720 40 -35.00V 0.05U 2W 41.001 HEAD DC VOLTAGE TESTS: {400V Range} 41.002 5720 400 350.0V 0.5U 2W 42.001 HEAD DC VOLTAGE TESTS: {1000V Range} 42.002 5720 1000 1000V 2U 2W 43.001 HEAD {1ms MIN/MAX TESTS} 43.002 ASK- Q 43.003 5720 2V 60H S 2W 43.004 DISP Press the MIN MAX button. 43.004 DISP Press the BEEPER button to display maximum values. 43.005 ACC 2.828VP 75P% 7e-6U 43.006 MEMI Enter UUT reading in volts peak: 43.007 MEME 43.008 MEMC Vp 0.045U 44.001 ASK+ Q 44.002 5720 2V 60H S 2W 44.003 DISP Press the MIN MAX button to display minimum values. 44.004 ACC -2.828VP 75P% 7e-6U 44.005 MEMI Enter UUT reading in volts peak: 44.006 MEME 44.007 MEMC Vp 0.045U 45.001 HEAD {DC MILLIVOLT TEST} 45.002 DISP Rotate the UUT function switch to mVDC. 45.003 HEAD DC MILLIVOLT TEST: {400mV Range} 45.004 5720 400 350.0mV 0.5U 2W 46.001 HEAD {RESISTANCE TESTS} 46.002 DISP Rotate the UUT function switch to Ohms. 46.003 HEAD RESISTANCE TESTS: {400 Ohm Range} 46.004 5720 400 190.0Z 0.5U CW 47.001 HEAD RESISTANCE TESTS: {40 kOhm Range} 47.002 5720 40 19.00kZ 0.05U CW 48.001 HEAD RESISTANCE TESTS: {4 MOhm Range} 48.002 5720 4 1.900MZ 0.005U 2W 49.001 HEAD RESISTANCE TESTS: {40 MOhm Range} 49.002 5720 40 19.00MZ 0.22U 2W 50.001 HEAD {CONDUCTANCE TEST} 50.002 DISP Press the Range button twice for 40nS range. 50.003 HEAD CONDUCTANCE TEST: {40nS Range} 50.004 5720 40 10.00nY 0.20U 2W 51.001 HEAD {DIODE TEST} 51.002 DISP Rotate the UUT function switch to Diode. 51.003 5720 3.000V 0.061U 2W 52.001 CALL Sub Fluke 83,85,87 (1 year) Capacitance CAL VER 52.002 JMP 53.001 52.003 EVAL dummy 53.001 HEAD {DC MILLIAMP TESTS} 53.002 DISP Rotate the UUT function switch to mA/A. 53.003 DISP Connect the 5720A and the UUT as follows. 53.003 DISP [32] 5720A OUTPUT HI to UUT mA/uA 53.003 DISP [32] 5720A OUTPUT LO to UUT COM 53.004 HEAD DC MILLIAMP TESTS: {40mA Range} 53.005 5720 40 35.00mA 0.09U 2W 54.001 HEAD DC MILLIAMP TESTS: {400mA Range} 54.002 5720 400 350.0mA 0.9U 2W 55.001 HEAD {AC MILLIAMP TESTS} 55.002 DISP Press the BLUE button to select AC. 55.003 HEAD AC MILLIAMP TESTS: {40mA Range} 55.004 5720 40 35.00mA 0.37U 60H 2W 56.001 5720 40 35.00mA 0.37U 1kH 2W 57.001 HEAD AC MILLIAMP TESTS: {400mA Range} 57.002 5720 400 350.0mA 3.7U 60H 2W 58.001 5720 400 350.0mA 3.7U 1kH 2W 59.001 HEAD {DC MICROAMP TESTS} 59.002 DISP Rotate the UUT function switch to uA. 59.003 HEAD DC MICROAMP TESTS: {400A Range} 59.004 5720 400 350.0uA 1.0U 2W 60.001 HEAD DC MICROAMP TESTS: {4000A Range} 60.002 5720 4000 3500uA 9U 2W 61.001 HEAD {AC MICROAMP TESTS} 61.002 DISP Press the BLUE button to select AC. 61.003 HEAD AC MICROAMP TESTS: {400A Range} 61.004 5720 400 350.0uA 3.7U 60H 2W 62.001 5720 400 350.0uA 3.7U 1kH 2W 63.001 HEAD AC MICROAMP TESTS: {4000A Range} 63.002 5720 4000 3500uA 37U 60H 2W 64.001 5720 4000 3500uA 37U 1kH 2W 65.001 DISP Disconnect the 5720A from the UUT. 65.002 HEAD {DC AMP TESTS} 65.003 DISP Rotate the UUT function switch to mA/A. 65.004 DISP Connect the 5725A and UUT as follows: 65.004 DISP [32] 5725A CURRENT OUTPUT HI to UUT A 65.004 DISP [32] 5725A CURRENT OUTPUT LO to UUT COM 65.005 HEAD DC AMP TESTS: {4000mA Range} 65.006 5720 4000 3500mA 9U B1 2W 66.001 HEAD DC AMP TESTS: {10A Range} 66.002 5720 10 10.00A 0.04U B1 2W 67.001 HEAD {AC AMP TESTS} 67.002 DISP Press the BLUE button to select AC. 67.003 HEAD AC AMP TESTS: {4000mA Range} 67.004 5720 4000 3500mA 37U 60H B1 2W 68.001 5720 4000 3500mA 37U 1kH B1 2W 69.001 HEAD AC AMP TESTS: {10A Range} 69.002 5720 10 10.00A 0.12U 60H B1 2W 70.001 5720 10 10.00A 0.12U 1kH B1 2W #76.001 CALL Sub Fluke 83,85,87 (1 year) Frequency CAL VER /3325 71.001 END #! T.U.R.s less than 4.00: 5 #! T.U.R.s estimated using RANGE value: 0 #! T.U.R.s not calculated (ASK- U): 7 #! T.U.R.s not computable at compile time: 0 #! FOR JUSTIFICATION REFER TO COMMENTS FOLLOWING EACH TEST IN THIS LISTING.